Principle |
Zirconia electro-chemical cell |
Structure |
SS box, TB-IIP sensor, C-101 control unit,
needle valve, flow meter etc. |
Model |
TB-FI |
Pressurized sample |
TB-FI-P |
Sampling pump installed |
TB-FI-PC |
Sampling pump and charcoal filter installed |
TB-FI-PR |
Sampling pump and charcoal filter and mist-separator
installed |
Measuring range |
1ppm~100%O2 |
Display |
%:0~99.99%O2
ppm:0~9999ppmO2 |
Output |
D.C.4~20mA, 0~1V or 0~10V(isolated)
F.S. can be set within the range of the above-mentioned display freely.
RS232C |
Range change |
Auto/Manual |
Initial setting |
0~25%O2/0~1000ppmO2 |
Alarm |
Output :Hi/Lo, HHi/Hi or Lo/LLo(each A Point of
contact・LCD Display)
Failure:Disconnection of Heater, RTD(LCD Display) |
Linearity |
Large one either of less than ±1%FS or±1ppm |
Repeatability |
Large one either of less than ±1%FS or±1ppm |
Response |
90% reading 10sec.(Swinging to a high density side) |
Drift |
Less than ±2%FS/wk |
Piping |
Rc1/4 |
Sample Flow Rate |
0.2~2L/min |
Sample Temp. |
300°C MAX. |
Sample pressure |
Others 203kPa~1.33×10-1Pa(2atm~10-3Torr) |
Warm up time |
About 20 minutes |
AC Power |
AC85V~AC260V |
Accessory |
Power supply code |