Super high density Oxygen Analyzer |
FGISX |
What's FGISX
This is high density oxygen analyzer using limiting current method of Zirconia electrolyte. |
|
FEATURES
- It can measure over 99.9% Oxygen at high accuracy and stability.
- It does not need reference gas and not be influenced by environment.
- Simple structure, maintenance free.
- Power saving.
|
APPLICATIONS
Air separation plant, Oxygen filling station,Liquid Oxygen tank, Oxygen piping maintenance, Oxygen PSA Laboratory etc. |
SPECIFICATIONS
Handling |
Portable |
Measuring Sphere |
0 ~ 100% O2 |
Display |
00.00% ~ 100.00% O2 |
Range |
1Range (0 to 100% O2 or 90 to 100%O2) |
Output Signal |
D.C.4 ~ 20mA or D.C. 0 ~ 10V |
Alarm Output |
Hi/Lo 2points (Option) |
Serial Comm. |
RS232 or RS485 (Option) |
Linearity |
Less than ±1% O2 (0% ~ 90%O2), ±0.1 O2 (90%~100%O2) |
Repeatability |
Less than ±1% O2 (0% ~ 90%O2), ±0.1 O2 (90%~100%O2) |
Response |
90% reading in 10 sec.(Swinging to a high density side) |
Sample Condition |
Not include water, corrosive, halogen, combustible
Introduce (positive pressure) |
Sampling |
0.5 ~ 2.0L/min |
Flow rate |
50°C MAX. |
Sample Temp. |
INLET&OUTLET Rc1/4 |
Warm Up Time |
About 10 minutes |
Power Source |
AC100±10V 50/60Hz 50VA MAX. |
Ambient Temp. |
0 to 50°C |
Attachment |
Power cord 3m |
Option |
Signal & communication cable 3m |
*For the improvement, the specification and design may be changed without prior notice.
|